Anaesthesia machine and capnograph simulator
NAVLEBCSESet oxygen flow, the APL valve, soda lime and the breathing circuit, then watch gas flow, circuit pressure and a live capnogram. Work through classic faults: exhausted soda lime, stuck valves, leaks and rebreathing.
- 1You change
Set the oxygen flow, vaporiser, APL valve and soda lime, pick a rebreathing or non-rebreathing circuit, and set how the patient breathes.
- 2Inside the circuit
Fresh gas (green) flows to the patient; exhaled gas carrying CO2 (amber) goes through the valves, bag and soda lime. Click any part of the machine to learn its job.
- 3On the capnograph
The live capnogram, end-tidal and inspired CO2 and circuit pressure change at once. Faults show up as classic waveform shapes.
The machine: click any part
Swipe sideways to see the whole circuit.
Click a part (flowmeter, vaporiser, valves, bag, APL valve, soda lime) to see what it does and how it fails. Click a valve to make it stick open.
Where the CO2 goes
Every breath the patient adds CO2 to the circuit. It must leave through the soda lime (circle) or be washed out by fresh gas (non-rebreathing). Whatever is left is breathed back in.
Pick a problem
Thumbnails show the shape each problem gives on the capnograph. Mystery fault sets a hidden fault: read the monitor and the machine, then name it.
Mystery fault
Score 0 / 0Starting point
Flowmeter
Vaporiser
Valves
APL valve
Soda lime
Bag
Sampling adapter
Capnograph
A 20 kg dog breathing spontaneously on a circle system: oxygen 1 L/min, isoflurane 2%, APL valve open, fresh soda lime. EtCO2 sits near 40 mmHg and the inspired CO2 returns to zero between breaths. Change one control at a time and watch the capnogram and the circuit pressure.
Study notes
Reading the capnogram
The four phases of one breath
Time-based capnogram from a spontaneously breathing patient.
| Phase | What happens | Normal |
|---|---|---|
| Phase IInspiratory baseline | Start of expiration: dead-space gas with no CO2 leaves first | 0 mmHg |
| Phase IIExpiratory upstroke | Dead-space gas mixes with alveolar gas; CO2 rises fast | Steep |
| Phase IIIAlveolar plateau | Alveolar gas; the highest point at the end is the EtCO2 | 35–45 mmHg |
| Phase 0Inspiratory downstroke | Inspiration of fresh gas washes CO2 out of the sensor | Steep, back to 0 |
Patterns and what causes them
| Pattern | Common causes |
|---|---|
| Baseline above 0 (rebreathing)FiCO2 > 5 | Exhausted soda lime, stuck-open valve, too little fresh gas on a non-rebreathing circuit, excessive equipment dead space |
| Slow, sloping downstroke | Incompetent inspiratory valve (exhaled gas goes back up the inspiratory limb) |
| Shark fin (slow upstroke, no plateau) | Bronchospasm (feline asthma), kinked or narrow tube, secretions, obstructed filter |
| High EtCO2, normal shape> 45 > 60 | Hypoventilation (deep plane, opioids, obesity, positioning), rebreathing, hyperthermia, malignant hyperthermia, bicarbonate |
| Low EtCO2, normal shape< 35 | Hyperventilation (light plane, pain, overzealous IPPV), low cardiac output, hypothermia, sampling leak |
| Sudden fall to near 0 | Cardiac arrest, disconnection, extubation, complete obstruction, massive embolism |
| A few small waves, then flat | Oesophageal intubation |
| Irregular, low plateau | Leak at the sampling site or around an uncuffed tube |
The machine and the circuits
Numbers to know (typical)
Standard teaching values; texts quote slightly different ranges.
| Item | Typical value |
|---|---|
| Rebreathing (circle) fresh gas, maintenanceLow-flow; minimum about 0.5 to 1 L/min | 20–40 mL/kg/min |
| Circle fresh gas, induction or changing depth | 50–100 mL/kg/min |
| Non-rebreathing (Bain, Jackson-Rees, Mapleson) | 200–300 mL/kg/min |
| Use a non-rebreathing circuit below | about 3–5 kg |
| Oxygen consumption | 5–10 mL/kg/min |
| Tidal volume | 10–15 mL/kg |
| Reservoir bag sizeAbout 5 to 6 tidal volumes | about 60 mL/kg |
| Oxygen flush | 35–75 L/min |
| Peak pressure when bagging a dog or cat | 10–20 cmH2O |
| Normal circuit pressure, APL open | 0–2 cmH2O |
| Normal EtCO2Up to about 45–50 is accepted under anaesthesia; above 60 needs ventilation | 35–45 mmHg |
Parts, jobs and classic faults
| Part | Job | Classic fault |
|---|---|---|
| Flowmeter | Sets fresh gas flow; read the top of a bobbin, the centre of a ball | Flow too low for the circuit → rebreathing or empty bag |
| Precision vaporiser | Adds a set % of one agent; out of circuit, agent-specific | Wrong agent filled, tipped, empty, left on |
| Unidirectional valves | Keep gas moving one way round the circle | Disc stuck open (moisture) → rebreathing |
| Soda lime canister | Removes CO2; makes heat and water | Exhausted → rising FiCO2; colour can fade back on resting |
| Reservoir bag | Holds gas for each breath; lets you see breathing and bag the patient | Too small or empty; over-distended with a closed APL |
| APL (pop-off) valve | Vents excess gas to the scavenger | Left closed → pressure rises, barotrauma, low cardiac output |
| Scavenger | Carries waste gas out of the room | Blocked active system can pull gas from (or push into) the circuit |
Rebreathing vs non-rebreathing
| Feature | Circle (rebreathing) | Bain / Mapleson (non-rebreathing) |
|---|---|---|
| How CO2 leaves | Soda lime absorbs it | High fresh gas flow washes it out |
| Valves and absorber | Yes: two one-way valves, canister | No |
| Fresh gas flow | Low (cheaper, less pollution) | High |
| Resistance to breathing | Higher (valves, canister) | Low: best for small patients |
| Heat and humidity | Kept (warm, moist gas) | Lost (cold, dry gas) |
| Speed of changing depth | Slow at low flow | Fast: inspired = vaporiser setting |
| Usual patient | Over about 3–5 kg; horses and cattle on large-animal circles | Under about 3–5 kg: cats, toy breeds, exotics |
Memory hooks
- Baseline off the floor = rebreathing. Check the soda lime, the valves, then the fresh gas flow.
- No CO2, no trachea. A flat capnogram after intubation means oesophagus or disconnection until proven otherwise.
- A sudden fall to near zero with the tube in place is cardiac arrest until proven otherwise. During CPR, EtCO2 above about 15 mmHg suggests good compressions; a sudden jump signals return of circulation.
- Shark fin = slow emptying: bronchospasm or a kinked tube.
- APL open unless you are giving a breath. Close it, squeeze, open it again.
- Circle: 20–40 maintenance, 50–100 to change depth. Non-rebreathing: 200–300 (mL/kg/min).
- Fresh gas enters after the canister and before the inspiratory valve; the APL and bag sit on the expiratory side.
- EtCO2 tells you about three things: ventilation, circulation (CO2 delivery to the lungs) and the equipment.
Exam check
For exam study only. Simplified teaching models: do not use these tools to make decisions for real patients. Last reviewed November 2026.